Annex (2)
Updated on: 2007-11-25
To The Accreditation Certificate No. JAS Cal. - 001 Dated 2005-10-31
For Jordan National Metrology Institute (JNMI)
at the Royal Scientific Society/ Amman

 

Scope of Accreditation

Calibration of Conventional Mass, Weighing Instruments and Temperature


 

Measurand

Measuring Range
Measurement Conditions /
Procedures
 
Best Measurement Capability
Calibration Methods/ Standards

Conventional Mass

 (OIML recommendation R111, class E2)

1 mg, 2 mg, 5 mg

10 mg

20 mg

50 mg

100 mg

200 mg

500 mg

1 g

2 g

5 g

10 g

20 g

50 g

100 g

200 g

500 g

1 kg

2 kg

5 kg

10 kg

 

0,003 mg

0,003 mg

0,003 mg

0,004 mg

0,005 mg

0,006 mg

0,008 mg

0,010 mg

0,012 mg

0,016 mg

0,020 mg

0,025 mg

0,030 mg

0,05 mg

0,10 mg

0,25 mg

0,5 mg

1 mg

2,5 mg

5 mg

§  In-house Method JNMISMP 43
[Revision No.: (2); Date: 2007-05-01] – Calibration of Weights – Primary.

 

§  In-house Method JNMISMP 44
[Revision No.: (2);      Date: 2007-05-01] – Calibration of Weights – Secondary.

Conventional Mass

(OIML recommendation R111, class F1)

20 kg

 

30 mg

Tempreture

Standard platinum resistance thermometers

(Calibration at fix point temperatures of ITS-90)

0,01 °C

29,7646 °C

156,5985 °C

231,928 °C

419, 527 °C

660, 323 °C

961,78 °C

 

8 mK

10 mK

10 mK

10 mK

15 mK

20 mK

50 mK

§  In-house Method

   JNMISMP 41  
 
[Revision No.: (2); Date: 2007-08-01] – Calibration of Temperature Measuring Instruments and Sensors (TC's & RTD’s) using Fixed Points.

Base metal thermocouples

(Comparison method)

0 °C

to

400 °C

Furnace

0,8 K

§  In-house Method

   JNMISMP 42

[Revision No.: (2);    Date: 2007-08-01] – Calibration of Temperature Measuring Instruments and Sensors (TC's & RTD’s) by Comparison.

> 400 °C

to

1100 °C

Furnace

2,0 K

Balances

(On-site calibration)

Up to 230 kg

EURAMET/cg-18/v01 (previously EA 10/18

3 ·10-6

 

§   In-house Method JNMISMP 45

  [Revision No.: (2); Date:

2007-05-01] - Calibration

of Balances.

 

List of employees in the laboratory who are technically responsible for issuing the calibration certificates in the scope of accreditation:

1- Director of JNMI/ Eng. Fawaz M. Al-Labadi.

2- Head of Physical and Mechanical Metrology Division/ Eng. Loai M. Qdairat.

3- Head of Electrical Metrology Division/ Eng. Ibrahim M. Sehweil.

4- Calibration Engineer and Quality Officer/ Mustafa F. Flaifel.

5- Calibration Engineer/ Sukaina Deebajeh.

6- Calibration Engineer/ Mariam Bishtawi.

7- Calibration Engineer/ Manal Abu Khalaf.

8-Calibration Engineer/ Duaa’ Flaifel

9-Calibration Engineer/ Khaldoun Al-Jazzazi

10- Calibration Technician/ Raf'e A. Batayneh.

11- Calibration Technician/ Mohammad Y. Al-Faleh.

12- Calibration Technician/ Mahmood S. Sayyah. 

     13- Calibration Technician/ Tareq Mofadi.