Annex (1)
Updated on: 2007-11-25
To The Accreditation Certificate No. JAS Cal. - 001 Dated 2005-10-31
For Jordan National Metrology Institute (JNMI)
at the Royal Scientific Society/ Amman

 

Scope of Accreditation

Calibration of AC & DC Current, AC & DC Voltage and DC Resistance
 

Measurand

Measuring Range
Measurement Conditions /
Procedures
 
Best Measurement Capability
Calibration Methods/ Standards

DC voltage

 

sources

 

fixed values

100 mV

1 V

1.018 V

10 V

100 V

1000 V

 

3,6 · 10-6 · U*

3,4 · 10-6 · U

3,8 · 10-6 · U

3,4 · 10-6 · U

3,4 · 10-6 · U

3,6 · 10-6 · U

§  In-house method JNMISMP 48 [Revision No.:(1); Date: 2007-10-01] - Calibration of Fluke 5720 A Multi-Function Calibrator & DC Voltage Sources.

DC voltage

 

sources

  10 mV       to    <100 mV

    100 mV   to    <1 V

    1 V          to    <10 V

    10 V        to    <100 V

 100 V        to    <1000 V

 

2 · 10-6 · U + 8 µV

9 · 10-6 · U + 7 µV

12 · 10-6 · U + 4 µV

13· 10-6 · U + 50 µV

20 · 10-6 · U + 0,25 mV

§  In-house method JNMISMP 01 [Revision No.:(1); Date: 2007-08-01] - Calibration of DC Voltage Source.

DC voltage

 

measuring

 

Instruments

 

 

  10 mV     to    220 mV

   >220 mV  to    2.2 V

   >2.2 V      to    11 V

   > 11 V      to    22 V

   >22 V       to    220 V

   >220 V     to    1000 V

 

1 · 10-6 · U + 6 µV

4 · 10-6 · U + 5 µV

4 · 10-6 · U + 5 µV

4 · 10-6 · U + 5 µV

5 · 10-6 · U + 40 µV

7 · 10-6 · U + 0,4 mV

§  In-house method JNMISMP 02 [Revision No.:(1); Date: 2007-08-01] - Calibration of DC Voltage Meter.

DC current

 

sources

10 µA        to    <100 µA

  100 µA      to    <1 mA

1 mA        to    <10 mA

10 mA       to    <100 mA

  100 mA     to    <1 A

  >1 A         to      11 A

 

100 · 10-6 · I* + 0,006 µA

45 · 10-6 · I + 0,009 µA

45 · 10-6 · I + 0,07 µA

68 · 10-6 · I + 0,9 µA

0,16 · 10-3 · I + 15 µA

0,64 · 10-3 · I +0,26 mA

§  In-house method JNMISMP 03 [Revision No.:(1); Date: 2007-08-01] - Calibration of DC Current Source.

DC current

 

measuring

 

Instruments

   10 µA       to    <220 µA    

   220 µA     to    2.2 mA

  > 2.2 mA   to    22 mA

  > 22 mA    to    220 mA

  > 220 mA  to   2,2 A

  >2,2  A      to   11 A

 

0,10 · 10-3 · I + 0,006 µA

35 · 10-6 · I + 0,007 µA

35 · 10-6 · I + 0,04 µA

50 · 10-6 · I + 0,7 µA

80 · 10-6 · I + 12 µA

0,74 · 10-3 · I + 0,50 mA

§  In-house method JNMISMP 04 [Revision No.:(1); Date: 2007-08-01] - Calibration of DC Current Meter.

DC resistance

 

sources

1 W          to  

>10 W      to

>100 W    to

>1 kW      to

>10 kW    to

>100 kW  to

>1 MW     to

>10 MW   to

>100 MW to

10 W

100 W

1 kW

10 kW

100 kW

1 MW

10 MW

100 MW

1 GW

 

0,11 · 10-3 · R* + 0,08 mW

32 · 10-6 · R + 1,1 mW

21 · 10-6 · R + 1,1 mW

20 · 10-6 · R + 11 mW

21 · 10-6 · R + 0,10 W

30 · 10-6 · R + 4,2 W

95 · 10-6 · R + 0,11 kW

1,2 · 10-3 · R + 1,2 kW

10 · 10-3 · R + 10 kW

§  In-house method JNMISMP 05 [Revision No.:(1); Date: 2007-08-01] - Calibration of DC Resistance Source.

DC resistance

 

Sources

 

Fixed

 

values

1 W

1,9 W

10 W

100 W

1 kW

10 kW

19 kW

100 kW

1 MW

10 MW

19 MW

100 MW

1 GW

 

22 · 10-6 · R

16 · 10-6 · R

11 · 10-6 · R

8,2 · 10-6 · R

5,8 · 10-6 · R

1,3 · 10-6 · R

5,0 · 10-6 · R

7,0 · 10-6 · R

15  · 10-6 · R

0,10 · 10-3 · R

0,11 · 10-3 · R

0,56 · 10-3 · R

10 · 10-3 · R

 

§  In-house method JNMISMP 52 [Revision No.:(1); Date: 2007-10-01] – Calibration of DC Resistance Source, Fixed Value.

DC resistance

 

measuring

 

Instruments

1 W

 1,9 W

10 W

100 W

1 kW

 10 kW

19 kW

100 kW

1 MW

 

 

62 · 10-6 · R

36 · 10-6 · R

13 · 10-6 · R

10 · 10-6 · R

8,0 · 10-6 · R

4,0 · 10-6 · R

7,0 · 10-6 · R

6,4 · 10-6 · R

16  · 10-6 · R

 

§  In-house method JNMISMP 06 [Revision No.:(1); Date: 2007-08-01] – Calibration of DC Resistance Meter.

DC resistance

 

measuring

 

Instruments

1,9 MW

10 MW

19 MW

100 MW

1 GW

 

22 · 10-6 · R

41 · 10-6 · R

52 · 10-6 · R

0,11 · 10-3 · R

10 · 10-3 · R

§  In-house method JNMISMP 06 [Revision No.:(1); Date: 2007-08-01] – Calibration of DC Resistance Meter.

AC voltage

 

sources

 

 

10 mV       to

100 mV     to

1 V            to

10 V          to

100 V        to

 

<100 mV

<1 V

<10 V

<100 V

700 V

 

40 Hz to 1kHz

40 Hz to 1kHz

40 Hz to 1kHz

40 Hz to 1kHz

40 Hz to 1kHz

90 · 10-6 · U + 18 µV

0,11 · 10-3 · U + 30 µV

0,1 · 10-3 · U + 0,25 mV

0,24 · 10-3 · U + 2,4 mV

0,47· 10-3 · U + 24 mV

 

§  In-house method JNMISMP 07 [Revision No.:(1); Date: 2007-08-01] – Calibration of AC Voltage Source.

AC voltage

 

measuring

 

Instruments

2,2 mV      to

22 mV       to

>220 mV   to

>2,2 V       to

>22 V        to

>220 V      to

22 mV

220 mV

2,2 V

22 V

220 V

1000 V

40 Hz to 1kHz

40 Hz to 1kHz

40 Hz to 1kHz

40 Hz to 1kHz

40 Hz to 1kHz

40 Hz to 1kHz

40 · 10-6 · U + 12 µV

80 · 10-6 · U + 13 µV

60 · 10-6 · U + 10 µV

60 · 10-6 · U + 43 µV

65 · 10-6 · U + 0,53 mV

80 · 10-6 · U + 3,5 mV

§  In-house method JNMISMP 08 [Revision No.:(1); Date: 2007-08-01] – Calibration of AC Voltage Meter.

AC current

 

sources

1 mA         to

1 mA         to

10 mA       to

10 mA       to

100 mA     to

100 mA     to

>1 A          to

 

 

<10 mA

<10 mA

<100 mA

<100 mA

1 A

1 A

11 A

40 Hz to 100 Hz

>100 Hz to 1 kHz

40 Hz to 100 Hz

>100 Hz to 1 kHz

40 Hz to 100 Hz

>100 Hz to 1 kHz

40 Hz to 60 Hz

0,7 · 10-3 · I + 2,4 µA

0,37 · 10-3 · I + 2,4 µA

0,71 · 10-3 · I + 24 µA

0,37 · 10-3 · I + 24 µA

1 · 10-3 · I + 0,24 mA

1,2 · 10-3 · I + 0,24 mA

0,60 · 10-3 · I + 1 mA

 

§  In-house method JNMISMP 09 [Revision No.:(1); Date: 2007-08-01] - Calibration of AC Current Sources.

AC current

 

measuring

 

Instruments

 

2,2 mA      to

 >22 mA    to

 >220 mA  to

 >2,2 A      to

22 mA

220 mA

2,2 A

11 A

40 Hz to 1 kHz

40 Hz to 1 kHz

40 Hz to 1 kHz

40 Hz to 1 kHz

0,14 · 10-3 · I + 0,3 µA

0,14 · 10-3 · I + 2,5 µA

0,28 · 10-3 · I + 35 µA

0,76 · 10-3 · I + 1 mA

§  In-house method JNMISMP 10 [Revision No.:(1); Date: 2007-08-01] - Calibration of AC Current Meters.

* U, I, R  = measured value

 

List of employees in the laboratory who are technically responsible for issuing the calibration certificates in the scope of accreditation:

1- Director of JNMI/ Eng. Fawaz M. Al-Labadi.

2- Head of Physical and Mechanical Metrology Division/ Eng. Loai M. Qdairat.

3- Head of Electrical Metrology Division/ Eng. Ibrahim M. Sehweil.

4- Calibration Engineer and Quality Officer/ Mustafa F. Flaifel.

5- Calibration Engineer/ Sukaina Deebajeh.

6- Calibration Engineer/ Mariam Bishtawi.

7- Calibration Engineer/ Manal Abu Khalaf.

8-Calibration Engineer/ Duaa’ Flaifel

9-Calibration Engineer/ Khaldoun Al-Jazzazi

10- Calibration Technician/ Raf'e A. Batayneh.

11- Calibration Technician/ Mohammad Y. Al-Faleh.

12- Calibration Technician/ Mahmood S. Sayyah. 

     13- Calibration Technician/ Tareq Mofadi.